Interpretation of Results
If the range rmax – rmin of the 25 radial distances is no greater than the manufacturers stated value of R then the performance of the probing system is verified. If the range exceeds the manufacturer’s stated value of R … Continue reading
Posted in Probing System Test 'R'
Leave a comment
Acceptance Test procedure
The user can choose the configuration of the stylus components of the probe but this must be within the limits specified by the manufacturer. Â It is recommended that the orientation of the stylus should not be parallel to any … Continue reading
Posted in Probing System Test 'R'
Leave a comment
Acceptance Test of the CMM Probe System
This test is used to establish whether the CMM is capable of measuring within the manufacturers stated value of R by determining the range of values of the radial distance r when measuring a reference sphere. It is advisable to … Continue reading