Category Archives: Probing System Test ‘R’

Interpretation of Results

If the range rmax – rmin of the 25 radial distances is no greater than the manufacturers stated value of R then the performance of the probing system is verified. If the range exceeds the manufacturer’s stated value of R … Continue reading

Posted in Probing System Test 'R' | Leave a comment

Acceptance Test procedure

The user can choose the configuration of the stylus components of the probe but this must be within the limits specified by the manufacturer.   It is recommended that the orientation of the stylus should not be parallel to any … Continue reading

Posted in Probing System Test 'R' | Leave a comment

Acceptance Test of the CMM Probe System

This test is used to establish whether the CMM is capable of measuring within the manufacturers stated value of R by determining the range of values of the radial distance r when measuring a reference sphere. It is advisable to … Continue reading

Posted in Probing System Test 'R', Test Artifacts | Leave a comment